Zero Calibrator
Matrix-matched material containing no measurable target analyte and used to establish the baseline of the calibration system.
A zero calibrator is a matrix-matched material containing no measurable concentration of the target analyte. It establishes the baseline response of a quantitative assay while closely resembling the biological characteristics of patient or research samples.
Unlike a simple solvent or reagent blank, a zero calibrator helps make the calibration system more representative of the actual sample matrix.
Biological specimens contain proteins, phospholipids, salts, metabolites, lipids, medications, and other endogenous components that can influence extraction, chromatography, ionization, binding behavior, and analytical response.
These materials may all contain little or no target analyte, but they serve different analytical purposes.
Matrix-matched material containing no measurable target analyte and used to establish the baseline of the calibration system.
Typically contains solvents or reagents without biological matrix and can help identify background from reagents or the analytical system.
Used to adjust concentration or prepare standards and may or may not resemble the intended biological specimen.
A matrix-matched zero calibrator helps establish the assay baseline before known calibrator concentrations are used to generate the calibration curve.
The biological specimen contains the target analyte within a complex sample matrix.
A matrix-matched material establishes the response when no measurable target analyte is present.
Known calibrator concentrations define the relationship between analytical response and concentration.
The sample response is compared with the calibration relationship to determine analyte concentration.