More Than 30 Minutes Before Reading the Plate

Reviewed by , Director of R&D  ·  Monobind Technical Support  ·  Updated

What if more than 30 minutes passed before I read the plate?

Plates should be read within 30 minutes of adding stop solution. After that, the end product of the enzyme reaction may precipitate and cause errors, so the data cannot be used with confidence. Run the test again.

Possible causes and what to do

Possible causeWhat to do
Technician or instrument errorRun the test again.
End product of the enzyme reaction may precipitate and cause errorsRun the test again.

How can I prevent this in future runs?

Confirm the reader is on, warmed up, and set to the correct wavelength before you add stop solution.

See our good lab practices for stop solution, lab environment, and substrate handling.

Related problems

Still seeing this problem?

Send us your product, lot number, and plate data and our team will help.

This guidance supports general troubleshooting. Always follow the Instructions for Use (IFU) and Certificate of Analysis (CofA) for your specific Monobind test system and lot.