High Calibrator Absorbance Causes Overflow: Causes and Fixes

Reviewed by , Director of R&D  ·  Monobind Technical Support  ·  Updated

Why does my highest calibrator overflow the plate reader?

Very high absorbance in the top calibrator usually means the room was warmer than 30 °C, the plate was rotated during incubation, or an incubation ran longer than specified. Higher absorbance does not harm the test itself, but results may be limited by your reader.

Possible causes and what to do

Possible causeWhat to do
Room above 30 °C, or plate rotated during incubationDo not shake, rotate, or heat the plate during incubation. You may measure absorbance at 405 nm, which has a lower extinction constant.
Substrate or reagent incubation longer than specifiedYou may measure absorbance at 405 nm, which has a lower extinction constant.

How can I prevent this in future runs?

Time each incubation precisely, keep the room at 20–25 °C, and leave the plate still on a level surface during incubation.

See our good lab practices for lab environment, substrate handling, and stop solution.