  • Controls Outside Expected Range: Causes and Fixes | Monobind

Troubleshooting Guide  ›  Assay Results

Controls Outside Expected Range: Causes and Fixes

Reviewed by , Director of R&D  ·  Monobind Technical Support  ·  Updated

Why are my ELISA controls outside the expected range?

Controls usually fall outside their range because the controls or calibrators are contaminated, or because the expected values being used are out of date. Re-run with fresh controls and calibrators and check the current values for your lot.

Possible causes and what to do

Possible causeWhat to do
Contamination of controlsRun the test again with fresh controls.
Contamination of calibratorsRun the test again with new calibrators.
Incorrect control valuesCheck the Monobind website for the updated control values for your lot.

How can I prevent this in future runs?

Use only the calibrators and controls listed on the Certificate of Analysis for your lot, change tips between each one, and keep vials capped.

See our good lab practices for lot-matched common reagents, micro-pipetting, and conjugate and tracer handling.

Related problems

Still seeing this problem?

Send us your product, lot number, and plate data and our team will help.

This guidance supports general troubleshooting. Always follow the Instructions for Use (IFU) and Certificate of Analysis (CofA) for your specific Monobind test system and lot.