  • Bad Duplicates in ELISA: Causes and Fixes | Monobind

Troubleshooting Guide  ›  Assay Results

Bad Duplicates in ELISA: Causes and Fixes

Reviewed by , Director of R&D  ·  Monobind Technical Support  ·  Updated

Why don't my ELISA duplicates agree?

Poor agreement between duplicate wells is usually caused by bubbles in the wells, a dispensing error, fingerprints or residue on the bottom of the plate, or an improper wash step. Correct the cause and read or re-run the plate.

Possible causes and what to do

Possible causeWhat to do
Bubbles in wellsUse a pin or needle to burst the bubbles.
Dispensing errorCheck the dispensing instrument.
Fingerprints or obstruction on the well bottomClean the bottom surface of the plate with DI water and dry it before measuring again.
Improper wash stepMake sure all wash equipment works properly and wells are filled completely but do not overflow.

How can I prevent this in future runs?

Pipette carefully down the side of the well, change tips between samples, handle strips by the edges, and check washer dispensing regularly.

See our good lab practices for micro-pipetting, washing, and microplate handling.

Related problems

Still seeing this problem?

Send us your product, lot number, and plate data and our team will help.

This guidance supports general troubleshooting. Always follow the Instructions for Use (IFU) and Certificate of Analysis (CofA) for your specific Monobind test system and lot.