  • Assay Drift in ELISA: Causes and Fixes | Monobind

Troubleshooting Guide  ›  Assay Results

Assay Drift in ELISA: Causes and Fixes

Reviewed by , Director of R&D  ·  Monobind Technical Support  ·  Updated

What causes drift across an ELISA plate?

Drift, where results shift steadily from the first wells to the last, is usually caused by interruptions during set-up or by reagents and samples that are not at room temperature. Prepare everything first and dispense without pausing.

Possible causes and what to do

Possible causeWhat to do
Interruption(s) during assay set-upSet up the assay without interruptions and in the shortest time possible while keeping good technique. Prepare all reagents and samples before you begin.
Reagents or samples not at room temperatureDispense samples and reagents only when all are at room temperature, to prevent temperature differences across the plate.

How can I prevent this in future runs?

Bring everything to room temperature, lay out samples in order, and use a multichannel pipette to shorten dispensing time.

See our good lab practices for lab environment, micro-pipetting, and microplate handling.

Related problems

Still seeing this problem?

Send us your product, lot number, and plate data and our team will help.

This guidance supports general troubleshooting. Always follow the Instructions for Use (IFU) and Certificate of Analysis (CofA) for your specific Monobind test system and lot.