NEW PRODUCTS

LuMatic
New CLIA plate analyzer, our most affordable reader yet is a big stride towards CLIA in every lab.

Free Testosterone
Will be released before MEDICA 2010.

PRODUCT ENHANCEMENT

CEA-NG
Calibration improved to match values of large systems. The 5th calibrator in this assay is 100ng/ml compared to 50ng/ml in our standard CEA kit. Patient samples with higher values now better correlate with big systems. Is it time you consider switching?

Progesterone
Sensitivity increased with greater ease of use. Ready to use enzyme reagent is a boon.

Digoxin
Made into HRP System. HRP and strepxavidin plate give customers benefits like better thermal durability and stronger bond.

Rapid TSH
Smaller sample volume and twice as fast.

Autoplex
Integrated analysis with several CLIA readers. Now Lumax, LuMatic, Impulse2/3 completely integrated results & reporting on single software interface.

EDUCATION CENTER

Additional Info on Value Ranges
hGH: Expected range for men and women
cTnI: Best sample types and normal range
Cortisol: Expected range in morning & afternoon for men, women and children

Lab Tips
Stored Calibration Curves & Use of Adjusters (2PTs)

Control Ranges
Monobind reports on its control as well as several of BioRad's (4 different products), we plan to limit to the most essential and need your vote to do so.

VOTE HERE for the biorad series you need and tell us the controls you want from Monobind

ASK US A QUESTION

UPCOMING PRODUCTS

Free Testosterone, ELISA & CLIA: 4Q2010
B12, Anemia, ELISA & CLIA: 1Q2011
Folate, Anemia, ELISA & CLIA: 1Q2011

VOTE FOR OUR NEXT PRODUCTS

HEALTH NEWS

First diagnostic test for Major Depressive Disorder MDD developed by Monobind customer
LEARN MORE

CUSTOMER SPOTLIGHT

Our most remote installation? Find Lumax @ 20,000 ft / 6,000 meters in Indian Himalayas. Sales reps travel two-days to reach the lab. Now that's dedication!

Tell us about your most remote or unique installation and you may win 3-free kits! Contest ends Dec 31, 2010, winner announced in next newsletter in Jan 2011

TELL US YOUR STORY

EDUCATION CENTER - Additional Info on Value Ranges

hGH Expected Range View PDF

Regarding your ELISA kit for determination of Growth Hormone (1725-300), do you have expected values for children, women and men separately or only general expected (normal) values stated in the product insert?

We can only provide a copy of reference ranges from a large clinical laboratory below which although is not Monobind's products, we would expect comparable results.

Growth Hormone (ng/mL): 004275
Volume: 0.8 mL (0.3 mL minimum) Serum

AGEMALE/FEMALE
1 day5 - 53
1 week5 - 27
1 - 12 months2 - 10
1 year to AdultLess than 5

Further, reference range for adult males and females please see info from MAYO CLINIC.

CTnI Expected Range View PDF

What is the literature used to establish the reference range for Monobind's Troponin-I immunoassays? How was Monobind's Troponin-I normal range established?

Troponin-I values are different in plasma and serum. In addition, plasma sample may be influenced by the additives used. For example, heparin minimally affects the results but oxalate and EDTA have significant effect. A serum sample that has been quickly separated from red cells is preferred. Based on the clinical data gathered by Monobind in concordance with the published literature the following ranges have been assigned. These ranges should be used as guidelines only:

Adult (Normal) < 1.3 ng/ml

Although the magnitude of numbers agrees between the methods, we do not have the adequate validation to lower the cut-off.

Cortisol Expected Range View PDF

What are the values to Cortisol in the morning and the afternoon, and the levels for children?

Based on literature search the normal values for Cortisol is given below:

POPULATIONMORNINGAFTERNOON
Adult5 - 23 ug/dL3 - 13 ug/dL
Child3 - 21 ug/dL3 - 10 ug/dL
Newborn1 - 24 ug/dL 

Please note: Normal results may vary from lab to lab.

EDUCATION CENTER - Lab Tips

Stored calibration curves and the use of adjusters (2PTs) View PDF

The calibration curve is very important because it shows the trend of the reacting analyte and is used to interpolate the control and/or sample values. If the calibration curve is not accurate the corresponding values calculated from the calibration curve will not be accurate. It is very important to run all the calibrators in duplicate for the first run of the assay. Once a 'master curve' has been established it is not required to run all calibrators for further runs of the assay within a 30-day period of the same reagent lot.

Once a 'master curve' has been established in an analyzer it is greatly suggested to run adjusters; Adjusters are calibrators that are near the two ends of the calibration curve that represent the variation that can occur between different runs of the assay. Monobind commonly refers to this as 2PT calibration, and its analyzers are programmed for this situation with necessary data analytics.

When running an assay with no adjusters the user assumes that there is no change in the assay, including the environment. This is not the best choice when a test is run in ambient conditions and/or on a laboratory bench that may be subject to drafts and environmental disruptions that may not be detrimental to the test but will affect the interpolation if adjusters were not run to compensate for these occurrences.

Whether a 'master curve' is being established or adjusters are being used in a run of an assay, it is best to run duplicates of each until experience with the assay shows significant positive reproduction through low CV% (e.g. <8% on Cal B is a good level). This will allow for sufficient data to determine if any bad duplicates, outliers, or environmental effects occurred. Once the assay's performance and the dispensing technique have been established, single wells of the calibrators/adjusters may be employed.

In a clinical setting there are three options given in an analyzer: New Curve, 2 PT, or Stored Curve. Table 1 shows the data that would be received from an analyzer when each method is used. The patient samples will be calculated based on the calibration curve option chosen. The New Curve and Stored Curve option use the values of the 'master curve' to calculate the sample concentrations; The 2PT calibration option will use the values obtained by the adjusters to adjust the calculated sample values to reflect how the assay curve varied for that specific run of the test.

The comparative values of each method can be seen more clearly in Graph 1 and Graph 2. Graph 1 depicts the Adjusters in relation to the Master Curve; the difference in the values does not seem vary significant in value but when graphed it is more evident. Graph 2 shows the results of the interpolation of the sample values using each one of these methods. Again the absorbance values do not very drastically from the ones received when the Master Curve was obtained but from the assay-to-assay variation, using strictly the stored curve gives drastically differing concentrations; however, when adjusters were used to calculate sample values the concentrations were those very similar to the 'master curve'.

At first thought, the user might believe using the stored curve will save time and resources but hopefully the data presented here explains otherwise. Every assay is different and as such some sample concentrations may not change between tests but adjusters will show when and where variation has occurred. The best option is to run a new calibration curve each time but results obtained from a stored curve with adjusters (2PT) can be used with confidence.

Table 1: Sample Assay Results received from ELISA Analyzer.

CONC
(ng/ml)
Master Curve2 PT
(Adjusters)
Stored Curve
(NO Adjusters)
A0.0435  
B5.1900.1500 
C25.6230  
D501.1350  
E1001.72751.900 
F2002.2275  
S01 0.1095 ⇒ 2.25 ng/ml0.086 ⇒ 2.23 ng/ml0.86 ⇒ 1.45 ng/ml
S02 1.4975 ⇒ 80.59 ng/ml1.64 ⇒ 80.72 ng/ml1.64 ⇒ 92.62 ng/ml
S03 0.1045 ⇒ 2.08 ng/ml 0.082 ⇒ 2.06 ng/ml0.082 ⇒ 1.31 ng/ml
S04 0.7405 ⇒ 30.74 ng/ml 0.780 ⇒ 30.88 ng/ml0.78 ⇒ 32.67 ng/ml
S05 1.31 ⇒ 64.77 ng/ml 1.42 ⇒ 64.41 ng/ml1.42 ⇒ 74.05 ng/ml

Graph 1: Master Curve and Adjusters.

Graph 2: Sample Values Calculated using each calibration method.